Technical Specifications

  Click here to download the pdf brochure

 

Moisture analysis method

Thermal analysis followed by coulometric measurement

Sample heating

In the built-in stainless steel oven
- From room temperature up to 400°C
- Adjustable in increments of 1°C

Detection limit

1 µg of water

Reproducibility

± 2% of absolute water value measured

Measuring range

Up to approx. 40% water

Sample weight, average

25–2,000 mg

Display

ppm/% and µg water

Analysis time

Average: 10–30 min | adjustable in increments of 1 min–10 h

Operator guidance | Software

English, for Windows® 2000 | NT | XP

Data storage

On the hard drive of the interfaced PC

Number of measuring programs

Limited only by PC’s hard drive memory

Power supply

110 - 230V

Frequency

50 … 60 Hz

Carrier gas

Integrated air compressor and gas drying unit, 
Connection for externaly supply with Nitrogen N2 (class 5.0); 
Gas prepressure 1 bar, 
Consumption 100–200 ml/min

Power consumption

Standby 100 W | At full power 600 W

Dimensions (W + D + H)

500 + 500 + 180 mm

Weight

approx. 20 kg

 

Accessories:

  • Regeneration kit for the electrolytic cell
  • Calibration standard
  • Nickel scoops for weighing samples
  • Electrochemical cell, uncoated
  • Molecular sieve for drying unit
  • Temperature calibration unit for the oven

 

Editorial   © Sartorius AG 2007   www.sartorius.co.uk